The surgeon was using the endo stitch with v-loc 180 absorbable sutures.The endo stitch failed to work after two sutures.The surgeon felt the v-loc 180 absorbable sutures were positioned wrong from the manufacturer.The surgeon called for another endo stitch with v-loc 180 absorbable sutures.The second endo stitch failed to work.The surgeon felt the v-loc 180 absorbable sutures were positioned wrong from the manufacturer.The surgeon called for a third endo stitch with v-loc 180 absorbable sutures.On placement of the first suture the tip of the endo stitch broke off while inside the vaginal cuff tissue.The surgeon was unable to see the tip of the endo stitch.X-ray was performed to locate the tip of the endo stitch.Once located the vaginal tissue contained the tip of the endo stitch was excised.Decision was then made to complete the closure vaginally.Vicryl suture was used to close the vaginal cuff.A cystoscopy was then performed.Patient taken to recovery in stable condition.No untoward patient effects.
|