Model Number 2360-71-000
Device Problems Break (1069); Device-Device Incompatibility (2919); Material Deformation (2976); Naturally Worn (2988); Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/30/2022
Event Type  malfunction  
Manufacturer Narrative
Product complaint # pc-(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the threads for cup impactor handle are damaged and nothing will attach.Orange tip has begun to disassociate from metal.No negative effects on patient.No pieces broke off in patient and did not broke into more than one piece.
 
Manufacturer Narrative
Product complaint #
=
> pc-(b)(4).Investigation summary
=
> no device associated with this report was received for examination.The investigation could not verify or draw any conclusions about the root cause of the reported event without the device to examine.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
 
Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: the device associated with this report was returned for analysis.Visual examination of the device found that the threaded tip is slightly stripped and the overall all condition of the device denotes wear signs as well as scratches consistent with heavy use for a long period of time.However, no deformed nor breakage condition was observed on the device.Additionally, a functional test with the mating device pinn poly impactor tip 36mm (product code: 221750008) assembled and disassembled the duraloc str cup impactor (product code: 236071000) properly.Unable to assemble allegation cannot be confirmed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURALOC STR CUP IMPACTOR
Type of Device
HIP INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15000742
MDR Text Key297004144
Report Number1818910-2022-12902
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295119906
UDI-Public10603295119906
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2360-71-000
Device Catalogue Number236071000
Device Lot NumberAG0706
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-