MAUDE Adverse Event Report: RICHARD WOLF GMBH RIGID GRASPER FORCEPS; RIGID GRASP. FORCEPS 5FR WL 430MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RICHARD WOLF GMBH RIGID GRASPER FORCEPS; RIGID GRASP. FORCEPS 5FR WL 430MM Back to Search Results
Model Number 8642.65
Device Problems Break (1069); Material Fragmentation (1261)
Patient Problem Foreign Body In Patient (2687)
Event Date 04/28/2022
Event Type  malfunction  
Manufacturer Narrative
Rwmic is submitting this report on behalf of the manufacturer richard wolf gmbh.Rwgmbh considers this mdr open.Rwgmbh will submit a follow up report after the device evaluation has been completed and/or new information becomes available.
 
Event Description
Richard wolf medical instruments corp.(rwmic) reference complaint (b)(4).It was reported to rwmic through a mw5110124 report that "wolf 5fr rigid forcep broke during use while trying to retrieve a polyp.Operative area had to be searched and irrigated profusely to locate the tiny forcep pieces." during a hysteroscopy endometrial ablation, forceps used for tissue retrieval were noted to be broken while in endometrial cavity.X-ray done & missing piece was found and removed.Patient outcome: there was no patient impact or any after effects during the procedure.Rwmic mdr awareness date: 06/17/2022.Additional information provided by the initial reporter: will the device be returned? yes.Was the device being used on a patient when the reporting issue occurred? yes.Was there any injury or illness to the patient due to the reported issue? no.Was there any injury or illness to any other personnel due to the reported issue? no.Did the issue cause a delay in the procedure being performed? unknown.(however; there is documentation to suggest there was a delay).Did the delay put the patient at risk? yes.Was there a similar back-up device available for use? yes.Was the scheduled procedure completed? yes.The initial reporter also indicated that an "xr was called in to help locate the broken piece.".
 
Event Description
Richard wolf medical instruments corp.(rwmic) reference (b)(4).The purpose of this submission is to report the results of the device evaluation.See h10.
 
Manufacturer Narrative
Rwmic is submitting this report on behalf of the manufacturer richard wolf gmbh.Follow-up report #1 is to provide fda with the results of the device investigation (product history evaluation) regarding this mdr.New information: the following fields have new information: b4, b5, g6, h2, h3 (not returned to manufacturer), h6 (type of investigation, investigation findings, investigation conclusion), and h10.The results of the request for device return: on july 6, 2022, rwmic has requested the user facility for the return of the device.A repair return authorization document no.200421455, (form 3043 08) has been issued.Currently, rwmic has no record of device being returned.Since device was not returned, the manufacturer, rwgmbh, has performed and completed a product history evaluation in lieu of the actual device investigation.The results of the product history evaluation: (rwgmbh, investigation report product histroy evaluation, form 5253 (00).Dated, 10/25/2022.The reported device rigid grasp forceps 5fr wl 430mm 8642.65 from batch #4500170290 was produced on 06/18/2015 and it contains 43 pieces.All 43 pieces were sold to rwmic till 07/29/2015.Rwgmbh has no records about when the affected end user received the device in question.There is no other complaints from the same batch.Since the reported device won't be returned for investigation, the root cause of the malfunction cannot be determined.Based on the received information as well as on older complaints, the probable root cause of the reported issue is mechanical overload.The complaint database at rwgmbh has been reviewed between 01/01/2019 and 07/16/2022.Rwgmbh has recorded 3 complaints regarding the device 8642.65: 700017515 from belgium.700018299 / 22-00185 (current complaint) from the us.The device of complaint (b)(4) was returned for investigation.The root cause of the reported issue of complaint (b)(4) was mechanical overload.The devices of the following complaints were not returned for investigation: 70017391 / 21-00129.700018299 / 22-00185 (current complaint).The product history evaluations are available for these complaints.In the relevant ifu ga-e193 / usa / 2011-08 v1.0 / eco 2011-0319, the user is advised in section 5 checks about the control of the device for damage and completeness as well as in section 6 application about the limited strength of the device.In the relevant risk assessment p08fo015 risk assessment - b 1-2: reusable non-optical forceps and scissors, rev.R05, the possible risk due to failure of the corresponding extend of damage and the assumed probability of occurrence and probability of reoccurrence were considered and assessed as an acceptable risk.Rwgmbh considers this mdr closed.An additional report will be submitted if new information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIGID GRASPER FORCEPS
Type of Device
RIGID GRASP. FORCEPS 5FR WL 430MM
Manufacturer (Section D)
RICHARD WOLF GMBH
pforzheimer strasse 32
75438 knittlingen, germany
GM 
Manufacturer (Section G)
RICHARD WOLF GMBH
pforzheimer strasse 32
d-75438 knittlingen, germany
GM  
Manufacturer Contact
zsófia bartholome-nagy
pforzheimer strasse 32
d-75438 knittlingen, germany 
GM  
MDR Report Key15013830
MDR Text Key301941749
Report Number9611102-2022-00031
Device Sequence Number1
Product Code GEN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8642.65
Device Catalogue Number8642.65
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HTA DEVICE - WOLF SCOPE
Patient Outcome(s) Other; Required Intervention;
Patient Age41 YR
Patient SexFemale
Patient Weight105 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-