Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2022
Event Type  malfunction  
Event Description
It was reported by a field service representative (fsr) that a purge failure occurred during use on a patient.As a result, the pump was swapped out.There was no report of patient complications, serious injury, or death.Fsr was unable to reproduce symptom.
 
Manufacturer Narrative
Qn#(b)(4).
 
Manufacturer Narrative
(b)(4).The reported complaint of "purge failure alarm" is not able to be confirmed.No part or recorder strip was returned to manufacturer for investigation.The field service agent serviced the pump and could not reproduce the issue.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
 
Event Description
It was reported by a field service representative (fsr) that a purge failure occurred during use on a patient.As a result, the pump was swapped out.There was no report of patient complications, serious injury, or death.Fsr was unable to reproduce symptom.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT2 WAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
marling heaton
3015 carrington mill blvd
morrisville 27560
MDR Report Key15014819
MDR Text Key304715442
Report Number3010532612-2022-00254
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902051715
UDI-Public30801902051715
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN000320
Device Catalogue NumberIAP-0500
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
NONE REPORTED; NONE REPORTED
-
-