• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS VITAMIN D TOTAL II; VITAMIN D TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS VITAMIN D TOTAL II; VITAMIN D TEST SYSTEM Back to Search Results
Model Number VITD TOTAL
Device Problems High Test Results (2457); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/23/2022
Event Type  malfunction  
Event Description
There was an allegation of a questionable high vitamin d total elecsys result from an unknown roche analyzer.The initial result was 100 ng/ml and the repeat result using another instrument was 26 ng/ml.The questionable result was not reported outside of the laboratory.The reagent lot number was either 599988 or 602107.Clarification was requested but was not provided.The expiration date was requested but was not provided.
 
Manufacturer Narrative
The investigation is ongoing.
 
Manufacturer Narrative
The investigation did not identify a product problem.The cause of the event could not be determined.A general reagent issue was excluded.The reagent expiration date was provided as jan-2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS VITAMIN D TOTAL II
Type of Device
VITAMIN D TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15026962
MDR Text Key304722701
Report Number1823260-2022-02105
Device Sequence Number1
Product Code MRG
Combination Product (y/n)N
Reporter Country CodeGR
PMA/PMN Number
K113546
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVITD TOTAL
Device Catalogue Number07464215190
Device Lot Number60210700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-