Model Number AE-QAS-SP42
Device Problem Device Appears to Trigger Rejection (1524)
Patient Problem Neuropathy (1983)
Event Type  malfunction  
Event Description
It was reported that there was an issue with an unknown activl device via information from the 2020 ess surgeon survey; however a specific form was not provided.According to the complainant, a sensory neurological event was observed postoperatively at vertebrae l4-l5 and l5-s1.The initial surgery was total disc replacement (tdr) with implantation of spike endplate.The physician noted "transient numbness" and "paresthesias".There was temporary impairment.No further patient details are currently available.The adverse event is filed under aag reference (b)(4).
 
Manufacturer Narrative
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: due to the fact that a lot number was not provided, the device quality and manufacturing history records could not be checked.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based on the investigation results, a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLLECT.NO.QAS SPINE ANTERIOR STABILIS.
Type of Device
SPINE SURGERY
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key15028671
MDR Text Key302246075
Report Number9610612-2022-00205
Device Sequence Number1
Product Code MAX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAE-QAS-SP42
Device Catalogue NumberAE-QAS-SP42
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-