MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number 5C4482
Device Problems Fluid/Blood Leak (1250); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter facility name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a minicap extended life pd transfer set had a "clicking sound" when the twist clamp was turned.This occurred during use of the device for peritoneal dialysis therapy.No disinfectant was used to disinfect the transfer set.The transfer set was in use for approximately 1.5 to 2 months.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
B5: upon additional information, fluid flow was observed with the clamp closed with the transfer set.The event occurred during an unknown process step.There was no report of patient injury or medical intervention associated with this event.The sample was received for evaluation.A visual inspection with the naked eye noted a broken occluder feet.Functional testing including leak, clear passage, clamp function and torque engagement testing were performed; leak and clamp function testing revealed a flow through the set.The reported condition was verified.Torque and clear passage testing were performed with no issues.The cause of the condition could not be determined; however, the damaged set is consistent with damage caused by exposure to chemical agents such as foreign solvents, dyes, or cleaning agents that damage the transfer set materials.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15042708
MDR Text Key297086966
Report Number1416980-2022-03686
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K152675
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C4482
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MINICAP
-
-