Model Number SDM-05000-2D2
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2022
Event Type  malfunction  
Event Description
It was reported that after switch panel recent replacment c-arm uncommanded rotation was noticed.No injuries were reported.Field engineer was dispatched and after investigation c-arm switches were replaced.Once this was completed the system was left working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM
Type of Device
FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC.
36 & 37 apple ridge road
danbury CT 06810
Manufacturer (Section G)
HOLOGIC, INC.
36 & 37 apple ridge road
danbury CT 06810
Manufacturer Contact
evelyn lopez
36 & 37 apple ridge road
danbury, CT 06810
MDR Report Key15044852
MDR Text Key301936237
Report Number1220984-2022-00061
Device Sequence Number1
Product Code MUE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSDM-05000-2D2
Device Catalogue NumberSDM-05000-2D2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse