Device evaluation of monitor has been completed.The reported problem (does not communicate with belt) has been confirmed.Upon investigation the monitor was not properly producing a driven ground signal.The cause for the failure was isolated to a defective r781 driven ground resistor on the computer/analog board.After replacement ran autotest without any issues.The root cause for the open resistor could not be positively identified.There was no adverse event that resulted from the damaged monitor.
|