• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. NATURAL TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING LEFT SIZE F; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. NATURAL TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING LEFT SIZE F; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/10/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Visual examination of the returned product identified the inner blister was not completely sealed as the adhesive residue is not uniform and does not meet the minimum requirements.The device history record was reviewed and no discrepancies relevant to the reported event were found.The likely condition of the product when it left zimmer biomet is non-conforming.However, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that when the circulator was opening the outer sealed package she stated it came off way too easily.It was determined that in order to be safe she would not pass it to the surgical field.She proceeded to open and pass the inner sealed package.There was no consequences or impact to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATURAL TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING LEFT SIZE F
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15064195
MDR Text Key304190995
Report Number0001822565-2022-02165
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024510579
UDI-Public(01)00889024510579(17)320319(17)65314702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173417
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42530007501
Device Lot Number65314702
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age48 YR
Patient SexFemale
-
-