MAUDE Adverse Event Report: CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE Back to Search Results
Model Number MCH-1000(I)
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2022
Event Type  malfunction  
Event Description
The customer reported that this device produced smoke during a cleaning procedure after the customer drained the device of water.They said that they do not know whether or not the device was finished with it's priming cycle when they started draining it, and they reported that the refrigeration module was not in the tank at the time that this occurred.
 
Manufacturer Narrative
Cardioquip's investigation determined the water level sensor was incorrectly installed.After all damaged components were replaced, the device passed final inspection and is fully operational.
 
Manufacturer Narrative
Further review determined that this is a non-reportable event.
 
Event Description
The customer reported that this device produced smoke during a cleaning procedure after the customer drained the device of water.They said that they do not know whether or not the device was finished with it's priming cycle when they started draining it, and they reported that the refrigeration module was not in the tank at the time that this occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOQUIP MODULAR COOLER HEATER
Type of Device
CARDIOPULMONARY BYPASS DEVICE
Manufacturer (Section D)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer (Section G)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer Contact
charley ford
8422 calibration ct.
college station, TX 77845
9796910202
MDR Report Key15084381
MDR Text Key304297913
Report Number3007899424-2022-00082
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMCH-1000(I)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-