Model Number MCH-1000(I) |
Device Problems
Fluid/Blood Leak (1250); Smoking (1585)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Cardioquip's investigation determined the water level sensor failed to activate causing the malfunction.After all damaged components were replaced, the device passed final inspection, and is fully operational.
|
|
Event Description
|
Customer complains of device smoking during cleaning procedure.
|
|
Event Description
|
Customer complains of device smoking during cleaning procedure.
|
|
Manufacturer Narrative
|
Further review determined that this is a non-reportable event.
|
|
Manufacturer Narrative
|
Further review determined that this is a non-reportable event.
|
|
Event Description
|
During their pm of the device, the technician had to fail the device due to smoke damage present which could not be resolved on-site.
|
|