The initial reporter received questionable elecsys ft4 assay results on a cobas 8000 - cobas e 602 module.The physician states that the ft4 results for one patient have not matched the clinical picture since 2016.The provided results are listed below by date.(b)(6) 2016: 3.68 ng/dl, (b)(6) 2016: 3.28 ng/dl, (b)(6) 2016: 4.25 ng/dl, (b)(6) 2018: 3.11 ng/dl, (b)(6) 2018: 2.52 ng/dl, (b)(6) 2019: 2.92 ng/dl, (b)(6) 2020: 3.29 ng/dl, (b)(6) 2021: 2.81 ng/dl, (b)(6) 2021: 3.30 ng/dl, the results were reported outside of the laboratory.The analyzer serial number is (b)(4).
|