The intellanav mifi open-irrigated ablation catheter was returned to boston scientific for analysis.Visual inspection revealed the presence of a foreign material in the distal tip and around the mini-electrodes.Functional testing noted no abnormal resistance while maneuvering the catheter.Continuity tests were performed and the results were within specifications.Pressure testing was performed to identify any potential leak paths and confirmed that the parameters were within specifications.Flow testing was performed, the catheter was attached to a pump to test the saline flow, the device resent a partial obstruction.Electrical testing was conducted and the device was able to complete a test ablation, but the noise values were found out of specifications.The results of the product analysis define the device pass the functional, leak, continuity tests, however in the electrical test, it was confirmed that the device had a significant noise and is out of the specifications.It was also noticed that the device had a foreign material in the irrigation holes and this issue could be affect the performance and capabilities of the device and cause the noise in the rhythmia mapping system.
|