• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problems Thermal Decomposition of Device (1071); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/21/2022
Event Type  malfunction  
Manufacturer Narrative
Section e : (b)(6) center.
 
Event Description
It was reported that no power was observed on the transmitter and scorch marks were observed on the transmitter's power adapter prongs.The transmitter with power adapter was replaced.There were no reported patient consequences.
 
Manufacturer Narrative
The field complaint of the transmitter not powering on was not verified.During the analysis, the transmitter could not be initially turned on as the alternating current (ac) power adapter had been cut off.The visual inspection revealed no damage to the outer plastic housing of the transmitter; however, the ac power adapter had been cut off and was no longer attached to the unit.- the unit was plugged into a working ac electrical wall outlet using ac power adapter tl# (b)(6).The ac power adapter had been cut off and removed.After opening the transmitter, no damages were observed on the main printed circuit board (pcb).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15106477
MDR Text Key296621338
Report Number2017865-2022-16115
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734504799
UDI-Public05414734504799
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Device Catalogue NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received02/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
-
-