• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING (PS) LEFT 13 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING (PS) LEFT 13 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problems Positioning Failure (1158); Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/07/2022
Event Type  malfunction  
Event Description
It was reported that an articular surface would not seat in the tibial tray.After cementing, the surgeon confirmed there were no substances such as soft tissue and that cement particles didn't remain between articular surface and tibia plate.After that he inserted the articular surface into tibia plate in accordance with the surgical technique, he found that there was 1mm or 2mm gap on anterior between these implants.He removed the surface and re-inserted it to the tibia plate.However, the same problem occurred again.Subsequently, 1mm thinner articular surface was used to complete the procedure.Attempts to obtain additional information have been made.However; no more information is available.
 
Manufacturer Narrative
(b)(4).Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual examination of returned product found it exhibits signs of use (nicked, gouged, impact marks) and the dovetail feature is flared.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE FIXED BEARING (PS) LEFT 13 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15111777
MDR Text Key303568063
Report Number3007963827-2022-00190
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024237278
UDI-Public(01)00889024237278(17)240930(10)64427535
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K121771
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42512400713
Device Lot Number64427535
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-