• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LIMITED BUBBLE CPAP SYSTEM; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LIMITED BUBBLE CPAP SYSTEM; BTT Back to Search Results
Model Number BC161-10
Device Problems Crack (1135); Loose or Intermittent Connection (1371); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A distributor in north carolina reported on behalf of a healthcare facility that the cpap probes of two bubble cpap generators were loose.The bubble cpap generator is part of the bc161-10 bubble cpap system kit.There was no reported patient involvement.
 
Manufacturer Narrative
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of investigation.
 
Manufacturer Narrative
Ps400391.Method: one of the two complaint bc161-10 bubble cpap system was returned to fisher & paykel healthcare (f&p) in new zealand, and it was visually inspected.Results: visual inspection of the complaint device revealed that the bubble generator was cracked from the top.Conclusion: we are unable to determine the cause of the reported event.However, the most likely cause of the reported event is physical damage due to impact.The existing design of the cpap probe is intended to be adjustable, which leads to the possibility of inadvertent movement.We note that there is a physical stop in the bubble cpap probe, to prevent the pressure from exceeding 10cmh20, and that a pressure manifold is used with the breathing circuit, to reduce the risk of unsafe circuit pressure.The user instructions that accompany the bubble cpap system kit illustrate in pictorial format the correct set-up and proper use of the bubble cpap generator.It also states the following: "always use pressure monitoring to verify that the patient is receiving the prescribed cpap level." "regularly observe the cpap generator for bubbling.If bubbling is not observed, check for and minimize the air leaks in the system and at the patient.If air leaks have been minimized, air flow may be increased to achieve continuous bubbling.".
 
Event Description
A distributor in north carolina reported on behalf of a healthcare facility that the cpap probes of two bubble cpap generators were loose.The bubble cpap generator is part of the bc161-10 bubble cpap system kit.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BUBBLE CPAP SYSTEM
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400, laguna canyon road
suite 300
irvine, CA 92618
9494534002
MDR Report Key15117586
MDR Text Key297439985
Report Number9611451-2022-00687
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBC161-10
Device Catalogue NumberBC161-10
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/12/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-