Model Number HL 20
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2022
Event Type  malfunction  
Manufacturer Narrative
It was reported that the hl 20 unit displayed "belt slip" error message during checking of the machine.No harm to any person was reported.A getinge field service technician (fst) was onsite and investigated the unit in question on 2022-07-16.The problem was resolved by replacing the optical tacho board.The system was checked according to factory¿s specification and all tests passed.The device was put back into use.Thus the reported failure "error message: beltslip" could be confirmed.The affected optical tacho board was not returned for further investigation, therefore no technical investigation could be performed.However the failure mode "belt slip" error message can be linked to the following most possible root causes according to the hl 20 risk management file: -defective/ dirty tacho, relay or pump belt a dirty foil or a defective optical tacho board is a plausible cause for the error message "belt slip".Because the sensor no longer detects one or more lines on the film, it measures an apparently low speed at the pump head.Since this speed is then lower than the motor speed, the pump interprets this as a slipping belt and displays this message.Device was manufactured in 2016-03-30.The review of the non-conformities during the period of (b)(6) 2016 to (b)(6) 2016 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported that the hl 20 unit displayed "belt slip" error message during checking of the machine.No harm to any person reported.Complaint id (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key15119348
MDR Text Key304801141
Report Number8010762-2022-00285
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number701043262
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-