LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Catalog Number 28-95-80 |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information were not provided.Livanova deutschland manufactures the s5 roller pump.The incident occurred in (b)(6).Part replaced: 90-305-770.Controlador de motor hms 0409.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that, during the procedure, the arterial s5 mast roller pump stopped.The control panel went black.The perfusionist started handcranking and he restarted the control panel.The screen worked again and returned an error on the s5 console, the pump didn't turn.The perfusionist switched the loop from the arterial pump to a backup pump.He reconfigured the alarms to the new pump.According to information , there was no impact on the patient.After the procedure, the perfusionist rebooted the pump, the arterial pump was working fine again.
|
|
Manufacturer Narrative
|
H.10: livanova field service technician was dispatched to customer facility and confirmed the reported issue.The motor control board was replaced, the unit was then tested and put back in service.According to complaints database analysis, no further complaint has been submitted for this unit since its installation in 2014.The root cause of the reported issue is traced back to a defective motor control board.
|
|
Event Description
|
See initial report.
|
|