LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number 28-95-80
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2022
Event Type  malfunction  
Manufacturer Narrative
Patient information were not provided.Livanova deutschland manufactures the s5 roller pump.The incident occurred in (b)(6).Part replaced: 90-305-770.Controlador de motor hms 0409.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that, during the procedure, the arterial s5 mast roller pump stopped.The control panel went black.The perfusionist started handcranking and he restarted the control panel.The screen worked again and returned an error on the s5 console, the pump didn't turn.The perfusionist switched the loop from the arterial pump to a backup pump.He reconfigured the alarms to the new pump.According to information , there was no impact on the patient.After the procedure, the perfusionist rebooted the pump, the arterial pump was working fine again.
 
Manufacturer Narrative
H.10: livanova field service technician was dispatched to customer facility and confirmed the reported issue.The motor control board was replaced, the unit was then tested and put back in service.According to complaints database analysis, no further complaint has been submitted for this unit since its installation in 2014.The root cause of the reported issue is traced back to a defective motor control board.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 MAST ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key15119636
MDR Text Key298397243
Report Number9611109-2022-00369
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number28-95-80
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/13/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-