An event of the valve ¿falling apart¿ was reported.Investigation of the device upon return to abbott found that the white spool had disconnected from the valve holder.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization, which includes inspection of the holder and inspection of the final valve assembly.No potential handling issues during holder attachment, final inspection, final sterilization, and packaging were identified that could cause the spool to detach from the holder.The cause of the detached spool could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.Medical device problem code: code 1506 (product quality problem) was removed.
|