This device is not distributed in us so that unique identifier is blank.This device is not distributed in the usa, therefore the pma/510(k) number is not applicable.If additional information becomes available, a supplemental report will be filed with the new information.
|
Use h-s2518 ((b)(4)) product.This product was opened to stop bleeding during gastric esd, inserted into a scope, and then the electrode cord was connected to a high-frequency device (vio300d).The forceps of this product were placed open at the bleeding point and energized.Initially, it was able to be energized, but since it was no longer energized in the middle, the use of this product was discontinued.There was no report of patient harm.The time of event is during use.
|
Correction information: g6: follow up #1, h2:if follow-up, what type? h3:device evaluated by manufacture, h6: coding changed based on the investigation result.Additional information: h4:device manufacture date.Evaluation summary: the following was confirmed about the device the device was considered to have been cleaned at a medical institution before being returned to us, so a small amount of blood, etc.Was adhered to the wire at the tip.When the control unit was slid, one cup opened and closed normally, but the other cup did not open and close because the link to the wire was disconnected.The device was connected to an inspection electrosurgical instrument (vio300d) to conduct an output test, and the cup was not energized because the link between the cup and the wire was disconnected.From the above, it was considered that when the product in question was temporarily removed from the endoscope during use and coagulated tissue or blood adhering to the cup was removed with gauze or the like, the gauze or the like caught on the wire, causing the wire to deform.When checking the opening and closing of the cup, the wire was gradually pulled toward the actuator and the link with the cup was disconnected, possibly resulting in an energization failure.
|