Model Number HL 20
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Customer reported that error message "head error" appeared on the roller pump module of the hl 20 machine during checking.No harm to any person was reported.A getinge field service technician (fst) was onsite and investigated the unit in question.The hl 20 machine was checked and it was found that the optical tacho board is defective and needs to be replaced.Thus the reported failure could be confirmed.The reported failure "head error" has been investigated in a previous complaint on (b)(6) 2019.The most probable root cause could be determined as a broken wiring of the optical tachoboard.The device in question was manufactured on 2013-05-16.The review of the non-conformities during the period of 2013-05-16 to 2022-07-28 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Customer reported that error message "head error" appeared on the roller pump module of the hl 20 machine during checking.No harm to any person was reported.Complaint id (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key15127703
MDR Text Key304358377
Report Number8010762-2022-00290
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number701043262
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/16/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-