As reported, during the preparation of this pressure monitoring set, the arterial line was occluded.It was impossible to unclog manually by flushing the device with the blue snap tab.Therefore, it required the usage of a 2 ml syringe to unclog the device applying force.There was no allegation of patient injury.Patient demographics were requested but not available.Follow up has started for device return.
|
Additional information were provided.Flushing this pressure monitoring set was still very difficult despite using a 2ml syringe.Therefore, the pressure sensor and the patient's red line were changed by new ones from another kit.No patient has been in contact with this system.Event date was added in b3.On the other hand, the pressure monitoring set involved in this case was received by our product evaluation laboratory for a full evaluation.The reported event of occlusion was confirmed.Flow restriction was detected from the kit during priming and flushing.Occlusion was found at bond joint between pressure tubing and female connector which was connected to zero stopcock.Bonding material formed a blockage inside pressure tubing and restricted the flow during priming.Drawings for model t001657a instructed to bond pressure tubing with the connector with bonding solvent.The flush device of the kit functioned properly.No other visible damage was observed from the kit.The risk assessment has been updated based on product evaluation results.Difficulty flushing/priming may have several root causes, but the end result would be an inability to flush the device during setup or inability to obtain blood pressure readings during use.The users of the device are well trained to trouble shoot, and if required, the device can be exchanged with minimal delay in monitoring/treatment.Therefore, potential for death or serious injury is remote.As such, this event is no longer considered reportable and a corrected supplemental report is being submitted.
|