Updated fields: d4, d9, g3, g6, h2, h3, h6, h10 the mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: upon evaluation of the unit, the lock showed rotational and lateral movement, but a residue buildup was present.New components were added to replace worn internal parts.General maintenance and cleaning performed.Root cause: complaint confirmed via inspection of the unit.There was movement in the lock and a residue buildup was present.Unit required cleaning and replacement of worn internal parts due to routine use and wear.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|