• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION ALL-IN-ONE CONTAINER; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION ALL-IN-ONE CONTAINER; CONTAINER, I.V. Back to Search Results
Catalog Number HRB8102P
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone no.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the all-in-one container was broken ¿damaged¿.This was observed prior to use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d9, h3, h4, h6 and h10.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Under water leak testing was performed which revealed a leak in the body of the bag due to a hole.The reported condition was verified.The cause of the hole could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALL-IN-ONE CONTAINER
Type of Device
CONTAINER, I.V.
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CALI
calle 36 no. 2c-22
apartado aero 2446
cali
CO  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15139498
MDR Text Key298999479
Report Number1416980-2022-03941
Device Sequence Number1
Product Code KPE
UDI-Device Identifier07707141303054
UDI-Public(01)07707141303054(10)
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHRB8102P
Device Lot NumberSE22BM9
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-