The customer reports during a colonoscopy procedure using auxiliary water tubing attached to a flushing pump and a scope, somehow the tubing became disconnected from the auxiliary water inlet on the scope.The user did not notice this issue and the user kept on activating the flushing pump, leaking approximately one gallon of water onto the flooring.The nurse did not notice the wet floor, and fell on her tailbone.The nurse was taken to the emergency room for evaluation and x-rays were taken to ensure there were no fractures.The nurse missed work due to this occurrence.No further consequences to the nurse have been reported.The user finished up the colonoscopy case.There was no impact to the patient as a result of this occurrence.The suspect device was returned to olympus.The user's report of the pump blowing off the tubing was not confirmed during inspection/evaluation.
|