Model Number FGS-1000
Device Problem Break (1069)
Patient Problem Failure of Implant (1924)
Event Date 07/07/2022
Event Type  Injury  
Event Description
It was reported that on an unknown date a fibulink was implanted into a patient along with a syndesmotic screw.The screw as well as the suture in fibulink broke post-operatively.It was reported that no revision surgery is necessary.This report is for the fibulink implant.This is report 2 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A photo investigation was completed: visual analysis of the photo revealed that the fibulink® syndesmosis repair kit/ss was found broken, the broken fragment was visible in x-ray provided.No other problems were reported.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for fibulink® syndesmosis repair kit/ss.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBULINK® SYNDESMOSIS REPAIR KIT/SS
Type of Device
WASHER, BOLT NUT
Manufacturer (Section D)
MONUMENT DEPUY SYNTHES PRODUCTS INC
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15163964
MDR Text Key297235243
Report Number1719045-2022-00025
Device Sequence Number1
Product Code HTN
UDI-Device Identifier00863176000300
UDI-Public00863176000300
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162805
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFGS-1000
Device Catalogue NumberFGS-1000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK - SCREWS: TRAUMA
Patient Outcome(s) Required Intervention;
Patient Age28 YR
Patient Weight159 KG
-
-