A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by reviewing the error logs and was able to reproduce the error by turning on the analyzer.While troubleshooting, error message 2062 tip remove appeared.Fse found many tips and removed them from the reagent tray and loader area.The fse pushed down the tip rack to release it from being stuck which was making the grinding noise.Fse resolved the complaint by moving the detach tip plate, releasing the tip rack, and removing the tips.Fse repaired and validated the analyzer by successfully performing quality control run without error and within acceptable range.No further action required by field service.The aia-900 analyzer is functioning as expected.(b)(4).The aia-900 operator's manual under section 12: error messages states the following: (4233) reag/tip-x home overrun cause: the home sensor s090, which is not supposed to be activated after the reagent loader moves, was activated.A retry will take place, and if there is no improvement a mf flag will be attached to the measurement result.Action: please contact tosoh local representatives.Check s090 and also check to see the cause of slipping, and so on, that occurs when pm090 moves to the limit side.(2062) tip removal error cause: a tip was detected during the tip removal check.A retry will take place, and if there is no improvement a mf flag will be attached to the measurement result.Action: if the trouble reoccurs, contact the tosoh local representatives.The most probable cause of the reported event was due to the tips on the reagent lanes and friction problem with the sampler.
|
A customer reported repeated error messages 4233 reagent/tip-x home overrun on the aia-900 analyzer.There was an audible grinding noise and the manual cover could not be opened.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for estradiol (e2), luteinizing hormone (lh ii), and prolactin (prl).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|