• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR RX ACCULINK CAROTID STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR RX ACCULINK CAROTID STENT SYSTEM Back to Search Results
Model Number 1011343-40
Device Problems Premature Activation (1484); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2022
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat a lesion in the carotid artery.The acculink self expanding stent system was inserted on the wire and during advancement, resistance was noted, so it was backloaded off the wire.The stent system did not enter the patient.Something did not feel right and it was noted that the stent was exposed, so it was not used.There was no adverse patient effect or a clinically significant delay in procedure.A same sized xact stent was implanted to complete the procedure.No additional information was provided.
 
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
The device was returned for analysis.The reported difficult to advance was unable to be confirmed.The reported premature activation was able to be confirmed.Additionally, multiple bends were noted on the shaft of the device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no reported damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that a coagulation of blood/contrast on the guide wire resulted in the reported difficult to advance.Manipulation of the device likely resulted in the noted multiple bends in the entire length of the shaft and ultimately resulted in the reported premature activation/stent deployment.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX ACCULINK CAROTID STENT SYSTEM
Type of Device
CAROTID STENT
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15188494
MDR Text Key302939624
Report Number2024168-2022-08613
Device Sequence Number1
Product Code NIM
UDI-Device Identifier08717648076374
UDI-Public08717648076374
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model Number1011343-40
Device Catalogue Number1011343-40
Device Lot Number1090961
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient Weight102 KG
-
-