(b)(4).Method: the complaint rt380 adult dual heated evaqua2 breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand for investigation, where it was visually inspected and analysed.Results: visual inspection identified a crack on the collar of the patient end of the expiratory limb.Further testing reveals environmental stress cracking due to chemical exposure on the collar.Conclusion: we are unable to determine the cause of the reported event.However, based on our investigation, it is likely due to exposure to chemicals causing environmental stress cracking.All rt380 adult dual-heated evaqua2 breathing circuits are visually inspected, and pressure and flow tested during production, and those that fail are rejected.The subject breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the rt380 adult evaqua2 breathing circuit show in pictorial format the correct set-up of the circuit and also states the following: - "check all connections are tight before use." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." - "do not soak, wash or sterilize this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers.".
|