Model Number MC2-5590R |
Device Problem
Migration (4003)
|
Patient Problems
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
Information was received that the right side rod has endcap separation.The patient is asymptomatic and the rod is still lengthening.No patient injury has been reported.
|
|
Manufacturer Narrative
|
The device has not been returned for evaluation as it remains in-situ.The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
The unit was not returned to nuvasive for direct investigation.During evaluation of the provided radiographic pictures it was noted that the threaded cap was unthreaded from the housing tube, therefore, the reported failure was confirmed.Additionally, the rod was observed to be distracted approximately beyond the midpoint.The issue recorded in this complaint is a known issue and a corrective action has been implemented.The device history records confirmed the unit was manufactured prior to the implementation of the corrective action.
|
|
Event Description
|
No additional information has been provided at this time.
|
|