NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM- MAGNETIC ACTUATION Back to Search Results
Model Number MC2-5590R
Device Problem Migration (4003)
Patient Problems Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Information was received that the right side rod has endcap separation.The patient is asymptomatic and the rod is still lengthening.No patient injury has been reported.
 
Manufacturer Narrative
The device has not been returned for evaluation as it remains in-situ.The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.
 
Manufacturer Narrative
The unit was not returned to nuvasive for direct investigation.During evaluation of the provided radiographic pictures it was noted that the threaded cap was unthreaded from the housing tube, therefore, the reported failure was confirmed.Additionally, the rod was observed to be distracted approximately beyond the midpoint.The issue recorded in this complaint is a known issue and a corrective action has been implemented.The device history records confirmed the unit was manufactured prior to the implementation of the corrective action.
 
Event Description
No additional information has been provided at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM- MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr, ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key15192417
MDR Text Key304446065
Report Number3006179046-2022-00238
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00812258022679
UDI-Public812258022679
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMC2-5590R
Device Lot Number9051501AAA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
Patient Weight28 KG
-
-