• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG CASPAR DISTR PIN16MMSTER; SPINE SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG CASPAR DISTR PIN16MMSTER; SPINE SURGERY Back to Search Results
Model Number FF905SB
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2022
Event Type  Injury  
Event Description
It was reported that there was an issue with ff905sb - caspar distr pin16mmster.According to the complaint description, the product "failed." this occurred during an anterior cervical fusion procedure.An additional medical intervention was required; broken pieces were retrieved.There was no surgical delay.Additional information was not provided.The adverse event / malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Manufacturer Narrative
Investigation results: there are no products available, therefore an investigation is not possible.Insufficient information and material to investigate.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Root cause cannot be finally concluded.Therefore the root cause specific risk cannot be identified.The potential risk determined during initial vigilance evaluation remains valid.Explanation and rationale: without the product, an exact cause cannot be determined at this moment.Due to the constant monitoring of the compliance with our quality standards, as matters stand, a production or material defect can most likely be excluded.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CASPAR DISTR PIN16MMSTER
Type of Device
SPINE SURGERY
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der gruen
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key15212178
MDR Text Key297764715
Report Number9610612-2022-00251
Device Sequence Number1
Product Code LXH
UDI-Device Identifier04038653364329
UDI-Public4038653364329
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFF905SB
Device Catalogue NumberFF905SB
Device Lot Number52552420
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-