MEDTRONIC HEART VALVES DIVISION EVOLUT PRO PLUS DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number D-EVPROP34 |
Device Problems
Material Separation (1562); Failure to Advance (2524)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during implant of this transcatheter bioprosthetic valve, there was difficulty while advancing the delivery catheter system (dcs) through the femoral artery.The dcs was unable to advance.The system was retrieved.Upon withdrawal, damage to the capsule was noted.A different dcs and valve were used to complete the procedure.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Additional information was received that there was also difficulty in loading the valve.Once the system was removed from the patient, a light thickness of the tip of the capsule, just before the nose cone, was observed.There was no injury to the femoral artery with the difficulty to advance the delivery catheter system (dcs).Per the physician, it was unknown as to what caused the capsule damage.No adverse patient effects were reported.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information was received which reported that while loading of the valve, in-habitual pressure was reported.The valve was loaded just once.The load was confirmed with visual, tactile and fluoroscopy methods.A misload was not reported.No additional adverse patient effects were reported.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Conclusion: the delivery catheter system (dcs) was not returned and images were not received.While loading of the valve, in-habitual pressure was reported.The valve was loaded just once.The load was confirmed with visual, tactile and fluoroscopy methods.Loading of the valve is a process in which the outcome is highly dependent on the operator experience and technique; the evolut instructions for use (ifu), contains instructions to use the integrated loading bath which features a mirror to ensure that all bioprosthesis outflow struts are symmetrical and captured in the capsule during loading.In addition, the ifu instructs to inspect the capsule visually and tactilely for a misloaded bioprosthesis.In this case the load was confirmed with visual, tactile and fluoroscopy methods.Difficulty while advancing the dcs through the femoral artery was reported.Difficulties advancing the dcs through the access vessel is known to be related to factors such as patient anatomy and physician technique, including guidewire and introducer sheath selection.In this case, a cause of the difficulties was unable to be determined given the information available.The dcs was unable to advance.The system was retrieved.Once the system was removed from the patient, a "light thickness of the tip of the capsule", just before the nose cone, was observed.With the information available and without the subject dcs for analysis or procedural images for review, a conclusive root cause of the reported event could not be determined.It is possible that the dcs caught on calcium in the anatomy during the first deployment attempt, however this could not be confirmed.A device history record (dhr) review was performed on the dcs and there were no correlations / issues identified regarding manufacturing.The device was manufactured per approved and released manufacturing processes and the device met all applicable manufacturing specifications prior to release for distribution.Updated data: h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|