FISHER & PAYKEL HEALTHCARE LTD ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT380
Device Problems Break (1069); Crack (1135); Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2022
Event Type  malfunction  
Event Description
A distributor reported on behalf of a healthcare facility in japan, via a fisher and paykel healthcare (f&p) field representative, that the expiratory limb temperature probe port of a rt380 adult dual-heated evaqua2 breathing circuit was found damaged.There was no reported patient consequence.
 
Manufacturer Narrative
(b)(4).The complaint rt380 adult dual-heated evaqua2 breathing circuit is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
 
Manufacturer Narrative
Ps(b)(4).Section d5 describe event or problem: a distributor reported on behalf of a healthcare facility in japan, via a fisher and paykel healthcare (f&p) field representative, that the collar of a rt380 adult dual-heated evaqua2 breathing circuit was found damaged.Method: the complaint rt380 adult dual-heated evaqua2 breathing circuit was received at fisher & paykel healthcare (f&p) in new zealand where it was visually inspected and analysed.Results: visual inspection of the returned rt380 adult dual heated evaqua2 breathing circuit confirmed that the patient end evaqua2 collar was cracked.Scanning electron microscopy (sem) analysis showed no indications of chemical exposure (environmental stress cracking) was detected.Further analysis of the returned complaint device revealed the surface features indicate that the failure mode is a brittle failure originating on the external surface of the collar.Conclusion: based on the analysis, the observed crack is most likely due to the collar being exposed to significant force.All rt380 adult dual heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject circuit would have met the required specifications at the time of production.The user instructions that accompany the rt380 adult dual-heated evaqua2 breathing circuit circuit states: - "do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
 
Event Description
A distributor reported on behalf of a healthcare facility in japan, via a fisher and paykel healthcare (f&p) field representative, that the collar of a rt380 adult dual-heated evaqua2 breathing circuit was found damaged.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key15213793
MDR Text Key297961259
Report Number9611451-2022-00735
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT380
Device Catalogue NumberRT380
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received10/12/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AERONEB NEBULIZER; AERONEB NEBULIZER; F&P MR850 RESPIRATORY HUMIDIFIER; F&P MR850 RESPIRATORY HUMIDIFIER; MUCOFILIN 20% (ACETYLCYSTEINE) INHALANT; MUCOFILIN 20% (ACETYLCYSTEINE) INHALANT; PB840 VENTILATOR; PB840 VENTILATOR
-
-