DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Fluid/Blood Leak (1250); Sparking (2595)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/11/2022 |
Event Type
malfunction
|
Event Description
|
It was reported the cart acted up during a case; it was leaking water, sparks flew out the bottom, and it shut off the solution panel in the or.The event occurred during surgery, however it was reported that there was no patient involvement.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the unit was confirmed to have a burnt power inlet module fuse.The leak and sparks could not be duplicated.The power supply was found with 120 v in but not 24 v out to cart board.The power inlet module and power supply were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed however, the sparks and leaking could not be duplicated.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information available.
|
|
Search Alerts/Recalls
|
|
|