Model Number N/A
Device Problems Fluid/Blood Leak (1250); Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2022
Event Type  malfunction  
Event Description
It was reported the cart acted up during a case; it was leaking water, sparks flew out the bottom, and it shut off the solution panel in the or.The event occurred during surgery, however it was reported that there was no patient involvement.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the unit was confirmed to have a burnt power inlet module fuse.The leak and sparks could not be duplicated.The power supply was found with 120 v in but not 24 v out to cart board.The power inlet module and power supply were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed however, the sparks and leaking could not be duplicated.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH SMOKE EVACUATOR
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15224035
MDR Text Key304922209
Report Number0001954182-2022-00077
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00889024466005
UDI-Public(01)00889024466005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00514010200
Device Lot Number0040157
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-