ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER LUXOR OPHTHALMIC MICROSCOPES; MICROSCOPE, OPERATING & ACC., AC-POWERED, OPHTHALMIC
|
Back to Search Results |
|
Model Number LX3 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A healthcare professional reported that an ophthalmic microscope head was found to be loose.The procedure details and patient impact were not reported.Additional information has been requested, but none received to date.
|
|
Manufacturer Narrative
|
The company representative confirmed and replicated the reported event.It was found that the that the upper arm, (which connects to the optical head), was moving without the brakes being disabled.To resolve the issue, the microscope was sent into service for repair.During the service repair the upper brake assembly was resecured and the microscope was reassembled.The unit was returned to the customer site.It was then tested and found to meet product specifications.A non-conformance-based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The root cause of the reported event is attributed to a loose upper brake assembly allowing the optical head to move freely with the arm.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|
|
|