U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER LUXOR OPHTHALMIC MICROSCOPES; MICROSCOPE, OPERATING & ACC., AC-POWERED, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER LUXOR OPHTHALMIC MICROSCOPES; MICROSCOPE, OPERATING & ACC., AC-POWERED, OPHTHALMIC Back to Search Results
Model Number LX3
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A healthcare professional reported that an ophthalmic microscope head was found to be loose.The procedure details and patient impact were not reported.Additional information has been requested, but none received to date.
 
Manufacturer Narrative
The company representative confirmed and replicated the reported event.It was found that the that the upper arm, (which connects to the optical head), was moving without the brakes being disabled.To resolve the issue, the microscope was sent into service for repair.During the service repair the upper brake assembly was resecured and the microscope was reassembled.The unit was returned to the customer site.It was then tested and found to meet product specifications.A non-conformance-based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The root cause of the reported event is attributed to a loose upper brake assembly allowing the optical head to move freely with the arm.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUXOR OPHTHALMIC MICROSCOPES
Type of Device
MICROSCOPE, OPERATING & ACC., AC-POWERED, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15231464
MDR Text Key297958121
Report Number2028159-2022-01189
Device Sequence Number1
Product Code HRM
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLX3
Device Catalogue Number8065753095
Device Lot Number13MTCN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/31/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-