• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3851
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the device was returned for evaluation.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 1mm distal of the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 10 atmospheres as per pcb2cm specification.A visual examination observed no damage to the tip or blades.All blades were present and fully bonded to the balloon surface.A visual and tactile examination identified no kinks or damage to the shaft of the device.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 28jul2022.It was reported that the blade was bent.The 95% stenosed target lesion was located in the severely tortuous and severely calcified left anterior descending artery.A 10mmx2.50mm wolverine coronary cutting balloon was selected for use.During the procedure, ablation was done with a rotapro 1.5 to 1.75mm and balloon was used for additional inflation.The balloon was then inflated five times at 12 atm but the blade was noted to be bent.All were removed from the patient's body and the procedure was completed with a different device.No patient complications were reported and patient was fine post procedure.However, device analysis revealed that the blade was lifted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15234358
MDR Text Key297979716
Report Number2124215-2022-30377
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0029383285
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-