MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA ARTICULAR SURFACE FIXED BEARING POSTERIOR STABILIZED (PS); PROSTHESIS KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA ARTICULAR SURFACE FIXED BEARING POSTERIOR STABILIZED (PS); PROSTHESIS KNEE Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/27/2022
Event Type  malfunction  
Event Description
It was reported that the device did not fit to mating component.Another device was used to complete the surgery.No further information is available.
 
Manufacturer Narrative
(b)(4).Patient identifier: (b)(6).Foreign source: singapore.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows damage to the distal and proximal surfaces in the form of nicks and gouges.The locking feature (dovetail) was found to flared out.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA ARTICULAR SURFACE FIXED BEARING POSTERIOR STABILIZED (PS)
Type of Device
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15236408
MDR Text Key304540981
Report Number3007963827-2022-00210
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024237087
UDI-Public(01)00889024237087(17)260507(10)65011325
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K121771
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42512400510
Device Lot Number65011325
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age82 YR
Patient SexFemale
-
-