Model Number OER-PRO |
Device Problems
Break (1069); Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/19/2022 |
Event Type
malfunction
|
Event Description
|
Olympus was informed by the biomedical engineer at the user facility that the customer automatic endoscope reprocessor ¿valve is leaking from a valve inside the unit¿.It was reported that the small valve below the yellow connector broke off.The customer problem was found during preventative maintenance being performed.No death, injury or harm was reported.
|
|
Manufacturer Narrative
|
The olympus field service engineer (fse) visited the user facility to perform an in-service on the subject machine.The fse confirmed the customer¿s problem and ordered the necessary parts.The biomedical engineer is trained on olympus endoscope reprocessors and will perform the repairs.The investigation is ongoing; therefore, the root cause of the reported phenomenon cannot be determined at this time.However, if additional information becomes available this medical device report will be supplemented accordingly.
|
|
Manufacturer Narrative
|
A review of the device history record found no deviations that could have caused or contributed to the reported issue.The cause of the reported event cannot be conclusively determined.Based on the results of the investigation, it is likely the connector could not be connected as some impact was added to the connector and the connector became loose and came apart.Probable causes for the event include: "* accumulated stress over time which caused the connector to get loose * unintentional impact to the connector with something hard." per the instructions for use: ¿check the following for each connector.The connector should be fixed firmly.The o-rings should be free of abnormalities such as cracks, tears, or dents.¿ and ¿warning -do not use the equipment if any connector seems to be damaged or defective.Using the equipment when an irregularity has been detected may interfere with reprocessing.Furthermore, fluid leakage may damage peripheral device or facilities near the equipment.¿ olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|