Model Number 3660
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 10/21/2021
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
It was reported that the patient's ipg was heating when therapy is on.As a result, the ipg was explanted and replaced.
 
Manufacturer Narrative
The reported issue was not able to be confirmed.Analysis of the returned ipg found it had no physical anomalies, it communicated with all lab utilities, and it passed all functional testing with no anomalous outputs observed.The returned device also met the product requirement specification for heat generation during use.During analysis no anomalies were identified that would have existed prior to explant that would have contributed to the alleged complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15246990
MDR Text Key298111503
Report Number3006705815-2022-16195
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067031419
UDI-Public05415067031419
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/27/2023
Device Model Number3660
Device Catalogue Number3660
Device Lot NumberA000116003
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS LEAD (X2)
Patient Outcome(s) Other;
Patient Age49 YR
Patient SexFemale
-
-