Model Number WA00014A
Device Problem Sparking (2595)
Patient Problem Skin Burning Sensation (4540)
Event Date 07/26/2022
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a therapeutic transurethral resection in saline (turis) procedure, the hf cable sparked causing minor burns to the surgeon's hands, which were treated with hand washing and disinfection.Reportedly, the output sound had been different than usual but the procedure was continued nevertheless until the sparks occurred.Then, the hf cable and main unit were replaced and the intended procedure was completed.
 
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation but to olympus medical system corporation (omsc), japan (returned to omsc on 2022-08-03).The evaluation at omsc confirmed that the hf-cable was broken at the area of the connector, where the voltage flashover injuring the doctor¿s hand had occurred.Age-related wear and tear in connection with a repeated tensile/bending load most likely caused single or all wires inside the cable to break.When the hf generator is activated, this can result in a voltage flashover in the damaged area.According to the article¿s lot number, the hf cable was manufactured in february 2018.It therefore must be assumed that the hf cable was used longer than the 12 months the cable is designed for.Thus, this event/incident can most likely be attributed to age-related wear and tear in connection with improper handling by the customer and thus to use error.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf cable without showing any abnormalities.The case will be closed on olympus side with no further actions.However, the reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-CABLE, BIPOLAR
Type of Device
HF-CABLES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
MED CONTACT GMBH
kornbühlstr. 100 ¿ 102
salmendingen
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key15247258
MDR Text Key301845481
Report Number9610773-2022-00332
Device Sequence Number1
Product Code FAS
UDI-Device Identifier04042761076449
UDI-Public04042761076449
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA00014A
Device Catalogue NumberWA00014A
Device Lot Number182W-2075
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OLYMPUS ELECTROSURGICAL GENERATOR "ESG-400".
Patient Age43 YR
Patient SexMale
-
-