MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD SYR ABG PRESET 3(1.6) LL 22X1 ECLIPSE BZ; SYRINGE WITH NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD SYR ABG PRESET 3(1.6) LL 22X1 ECLIPSE BZ; SYRINGE WITH NEEDLE Back to Search Results
Catalog Number 365061
Device Problem Filling Problem (1233)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/06/2022
Event Type  malfunction  
Event Description
It was reported when using the bd syr abg preset 3(1.6) ll 22x1 eclipse bz there was insufficient blood flow through the device.The following information was provided by the initial reporter.The customer stated: the syringe was without pressure when aspirating blood.Blood collection did not aspirate.".
 
Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by functional testing, each drawn with water, and no issues were observed relating to insufficient flow as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode insufficient blood flow.Bd was not able to identify a root cause for the indicated failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SYR ABG PRESET 3(1.6) LL 22X1 ECLIPSE BZ
Type of Device
SYRINGE WITH NEEDLE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15255450
MDR Text Key305245954
Report Number9617032-2022-00790
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2023
Device Catalogue Number365061
Device Lot Number1197994
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial