• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 - EUR2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 - EUR2 Back to Search Results
Model Number 27023
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/18/2022
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Performance testing confirmed the reported complaint.Review of the device data logs revealed battery charger fault alarms.The battery was replaced to address the issue.The device was serviced and fully tested before it was returned to the customer.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to an isolated component failure within the device battery assembly.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an internal battery degraded warning alarm.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - EUR2
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key15255594
MDR Text Key298206185
Report Number3007573469-2022-00547
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270231
UDI-Public(01)00619498270231(10)1414726
Combination Product (y/n)N
Reporter Country CodeFI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27023
Device Catalogue Number27023
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA08/18/2022
Distributor Facility Aware Date07/20/2022
Device Age32 MO
Date Report to Manufacturer08/18/2022
Date Manufacturer Received07/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-