• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM- MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM- MAGNETIC ACTUATION Back to Search Results
Model Number MC2-5070R
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Metal Related Pathology (4530)
Event Date 05/26/2022
Event Type  Injury  
Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.
 
Event Description
Information was received that there were signs of metallosis.The rod has been explanted and replaced with a traditional growing rod.Report 2 of 2.
 
Event Description
No additional information has been provided.
 
Manufacturer Narrative
The unit was not returned to nuvasive for investigation.During evaluation using the provided x-ray image it was noted that the rod had been partially distracted.There were no obvious device-related abnormalities noted.Based on the provided details there was insufficient information and no objective evidence found to indicate the presence of metallosis.The reported event is unable to be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM- MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr, ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key15259556
MDR Text Key298228963
Report Number3006179046-2022-00240
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00887517834010
UDI-Public887517834010
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberMC2-5070R
Device Lot Number7091514AAA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age8 YR
Patient SexFemale
-
-