Model Number CLV-190 |
Device Problems
Contamination (1120); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was evaluated by olympus.The evaluation found that the scope detection did not work and it was noted that humidity was present inside the channel pump.Additional evaluation findings are as follows: 1.) non-olympus lamp found installed in unit; 2.) output connector noted worn; 3.) dust accumulated inside the unit; 4.) pump unit failure; 5.) the output connector was found broken but the connection was possible.The investigation is ongoing and the definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
|
Event Description
|
The olympus, model clv-190, evis exera iii xenon light source was returned to olympus for repair due to a report that the electronic connector did not function.There was no patient involvement, associated with the event, reported to olympus.Upon inspection and testing of the returned device, it was found that the scope detection did not work and humidity was noted inside the channel pump.This report is submitted due to the findings that the scope detection did not work and humidity noted inside the channel pump, both identified during in-house service of the device.As the problems were found during in-house service of the device, there was no patient involvement.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 8 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, the phenomenon was confirmed.There was a trace of fluid inside the tube connected to the pump, and there was corrosion on the contact pins of the scope connector.Thus, the device might have been affected because it was connected with fluid after the scope was cleaned.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The replacement of pump, connector, lamp and cleaning are required to return the instrument to the oem specification.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|