Model Number CLV-190
Device Problems Contamination (1120); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated by olympus.The evaluation found that the scope detection did not work and it was noted that humidity was present inside the channel pump.Additional evaluation findings are as follows: 1.) non-olympus lamp found installed in unit; 2.) output connector noted worn; 3.) dust accumulated inside the unit; 4.) pump unit failure; 5.) the output connector was found broken but the connection was possible.The investigation is ongoing and the definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The olympus, model clv-190, evis exera iii xenon light source was returned to olympus for repair due to a report that the electronic connector did not function.There was no patient involvement, associated with the event, reported to olympus.Upon inspection and testing of the returned device, it was found that the scope detection did not work and humidity was noted inside the channel pump.This report is submitted due to the findings that the scope detection did not work and humidity noted inside the channel pump, both identified during in-house service of the device.As the problems were found during in-house service of the device, there was no patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 8 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, the phenomenon was confirmed.There was a trace of fluid inside the tube connected to the pump, and there was corrosion on the contact pins of the scope connector.Thus, the device might have been affected because it was connected with fluid after the scope was cleaned.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The replacement of pump, connector, lamp and cleaning are required to return the instrument to the oem specification.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15277603
MDR Text Key301681871
Report Number3002808148-2022-01295
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-