Model Number B35200 |
Device Problems
High impedance (1291); No Device Output (1435)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the patient was getting implant replacement and all the bipolar impedance looked good, but the monopolar impedances have open circuits.Left side showed bipolar impedances from 1200-2000 ohms, but monopolar is 6-7k ohms.The right side showed 5k ohms for 9 & 10, and monopolar for 8, 9, 10, 11 are out of range.Troubleshooting was not required.The manufacturer's representative (rep) said patient's impedance issue is chronic thus patient has not been able to turn therapy on.Rep didn't know when previous impedance issue started.Additional information was received from the manufacturer's representative (rep) who reported that the right side implant was the on ly side replaced and this was due to it nearing end of life and left side was not replaced.The rep stated the impedance issues were only on the left chest.The cause of the open circuits were not determined and further actions/interventions were not taken or planned.The issue was not resolved at this time.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|