Catalog Number 03P84-25
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/08/2022
Event Type  malfunction  
Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On 09-aug-2022, abbott point of care was contacted by a customer regarding i-stat crea cartridges that yielded suspected discrepant creatinine results on a patient.There was no patient information available at the time of this report.(b)(6).Test/collection times not provided.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The customer has not provided product lot information.The investigation is underway.
 
Manufacturer Narrative
Apoc incident: # (b)(4).The investigation was completed on (b)(6) 2022.A review of the device history record confirmed the cartridge lot passed finished goods release criteria.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.Ak, appendix 1- product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.
 
Event Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CREA CARTRIDGE
Type of Device
CREA+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key15279056
MDR Text Key305349263
Report Number2245578-2022-00112
Device Sequence Number1
Product Code CGL
UDI-Device Identifier10054749000125
UDI-Public10054749000125
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K973292
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03P84-25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/12/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-