MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/28/2022
Event Type  malfunction  
Event Description
The customer reported to olympus technical assistance center (tac), the endoscope reprocessor had a water flow issue.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
The subject device has not been returned to olympus for evaluation.During a follow up call with the customer, the customer noted the unit is now working fine.The olympus representative advised to verify when the last time the water filter was replaced.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Section g2 was corrected to accurately identify the report source as a healthcare professional.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the weaker water supply flow in the reprocessing bin could not be determined.It was observed that the primary side of the water filter was dirty due to a water issue that occurred in the area of the hospital.It is possible that the dirty water filter caused it to clog, which may have resulted in a weaker water flow.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15286941
MDR Text Key305252326
Report Number9610595-2022-01270
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/20/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-