Catalog Number E5MC4002 |
Device Problems
Filling Problem (1233); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/27/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the flow regulator of ten (10) uromatic tur administration sets were rigid which delayed the process of irrigation during cystoscopy.This issue occurred during patient use.The dripping of solution could not be observed; the tubing did not extend into the drip chamber.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H4: device manufacture date: september 2021.H10: ten (10) devices were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed including gravity and leak testing and no issues were observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|