Model Number D087031
Device Problems Fracture (1260); Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2022
Event Type  malfunction  
Event Description
During a persistent atrial fibrillation procedure, resistance was noted during insertion into the sheath and a fracture occurred approximately 1.5 cm from the tip.Another catheter was used to complete the procedure with no adverse consequences to the patient.
 
Manufacturer Narrative
One viewflex xtra ice catheter was received for evaluation.The catheter tip was noted to be bent and fractured.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The root cause of this issue was determined to be a design/process issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIEWFLEX¿ XTRA ICE CATHETER
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15296655
MDR Text Key305348494
Report Number3008452825-2022-00570
Device Sequence Number1
Product Code OBJ
UDI-Device Identifier05415067002082
UDI-Public05415067002082
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K133853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model NumberD087031
Device Catalogue NumberD087031
Device Lot Number8294873
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-